Scale behaviour

How the station turns load-cell readings into a weight: sampling, the stability rule, tare, calibration and the warnings it raises. The processing values are fixed in the firmware; only the load-cell profile, the calibration and the Spoolman tolerances are settings.

Steps for tare and calibration: Tare and calibrate.

Hardware

Item Value
ADC NAU7802 at I2C address 0x2A, SDA GPIO10, SCL GPIO11, 400 kHz
Sample rate 10 samples per second
Gain 128
Load-cell excitation (LDO) 3.0 V
I2C operation timeout 1 second
ADC missing or silent No sample for 1.5 seconds marks the scale disconnected; the station tries to start it again every 5 seconds.

At start-up the station scans the scale connector once and prints the result on the serial port. If it finds the ADC with the two lines swapped it prints NAU7802: FOUND WITH SDA/SCL REVERSED. Wiring: Wiring.

Load-cell profiles

Profile id Model Capacity Overload flagged above
yzc-133-5kg (default) YZC-133 5,000 g 5,500 g
yzc-133-2kg YZC-133 2,000 g 2,200 g

The overload threshold is capacity × scale_profile.overload_ratio (default 1.10). It is a warning only and does not protect the cell; never load a cell beyond its rated capacity. Changing the profile discards the calibration.

Sampling and stability

Step Rule
Filter The last 10 samples are sorted, the 2 lowest and 2 highest are dropped, and the rest are averaged.
Weight (filtered counts − zero offset) ÷ counts per gram
Stable The kept samples span at most 2.0 g, the reading is neither negative nor overloaded, and this holds for 1.5 seconds.
Weigh result Taken at the first stable moment and rounded to a whole gram. A steady reading within ±30 g of zero is reported as 0 g.
Timeout A measurement that is not stable after 20 seconds ends with an error.

Timeout messages:

Cause Message
Reading never settled "Scale is still moving. Leave the spool still and retry."
Overload "Scale overload detected. Remove weight and retry."
Steady reading below −30 g "Scale zero has shifted. Empty the platform, then use Tare in Settings > Calibrate scale and retry."
Calibration without a weight on the platform "No reference weight detected. Place the reference weight on the platform and retry."

Weighing requires a calibration ("scale calibration is required before weighing"). Only one measurement, tare or calibration runs at a time.

Tare

Tare stores the current filtered reading as the zero offset.

  • The raw reading must be steady: the kept samples of a full 10-sample window span at most 500 ADC counts. Otherwise: "scale must be stable before tare".
  • With a calibration present, the new zero is saved at once and survives a restart.
  • Tare is for the empty platform. The weight of an empty spool is handled separately, below.

Automatic zero tracking

Between measurements, when the platform reading is steady and within ±30 g of the stored zero for 3 seconds, the station moves a temporary zero correction toward the reading in steps of at most 5 g, up to 30 g in total. The correction is not saved; a tare, a calibration or a restart discards it.

Calibration

  1. Tare the empty platform.
  2. Place a known reference weight and start calibration with its weight in grams.
  3. The station waits until the reading has been steady for 4 seconds, then computes counts per gram = (filtered counts − zero offset) ÷ reference grams.
Rule Value
Reference weight More than 0 g, at most the profile capacity, at most 5,000 g over the API
Smallest accepted change 100 ADC counts and 5 counts per gram; less is treated as no weight placed
Stored values zero_offset_counts, counts_per_gram, reference_grams, load_cell_capacity_grams
Where In the configuration file and as a CRC-protected copy in NVS (key scaleCal)
Discarded by Changing the load-cell profile; factory reset
Kept by Restart; firmware update

Filament weight

measured filament = scale reading − empty-spool weight, computed only from a stable, non-negative reading.

The empty-spool weight is the first of these that is greater than 0:

  1. The spool's own empty weight in Spoolman (spool_weight). If the spool has none, Spoolman's filament value, then its vendor value, stands in here.
  2. The empty-container weight stored on the tag.
  3. The filament's spool_weight in Spoolman.
  4. The vendor's empty_spool_weight in Spoolman.

A value of exactly 0 is used only when no source holds a weight greater than 0. With no value at all the measurement cannot be saved: "The empty spool weight is unknown. Set it on the spool in Spoolman, then weigh again."

What happens to the measured value, including the 5 g tolerance and automatic saving: Spoolman integration.

Warnings

Flag Condition
Negative Weight below −2.0 g
Overload Weight above capacity × overload ratio, or the filtered ADC value at 98 % or more of its range
Drift A stable reading has moved more than 5.0 g from where it first became stable
Disconnected No sample for 1.5 seconds, or an I2C error

A negative or overloaded reading is never stable.

Status words on the touchscreen

Word Meaning
ERROR The ADC is not responding.
MEASURING A measurement started and fewer than 3 samples have arrived.
SETTLING A measurement is running and the reading is not yet stable.
STABLE The reading is stable, or the last measurement finished with a weight.
ZERO The last measurement finished at 0 g.
SETUP REQUIRED The scale has no calibration.
READY Calibrated, no measurement taken yet.